Hitachi Koki HITACHI
HomePower ToolsLife scienceFinancialCorprate ProfileContact us Japanese

Life-science > Products > Large-Scale Continuous Flow Ultracentrifuge > Rotors for CC40


Rotors for CC40/CC40S


Rotor Specifications for CC40
Rotor Specifications for CC40S

Rotors for CC40
Rotor model
C40CT4 *1
Type of cores Core (H)
(High Flow Type)
Ti Core (H)
(High Flow Type)
External appearance of cores *3
Max. speed (rpm) 48,000
Max. RCF (x g) 118,000
Max. revolving radius: Rmax (mm) 66
Rotor weight (kg) 31
Rotor material Titanium alloy
RCF (x g) at outside radius of center body 98,000
Radius of center body: Rmin (mm) 55
Core length (mm) 761
Core material Noryl® Titanium
Capacity (l) 3.2
Max. flow rate (l/hr) 50
O-ring material EPDM (Ethylene-propylene rubber, USP Class VI compliant)
Applications Density gradient centrifugation of virus, liposome, etc. (For high viscocity sample)

Rotor model
C40CT3
Type of cores Core (D) Ti Core (D) Core (E) *2 Core (F) Core (G)
External appearance of cores
*4

With preclarifier core
Max. speed (rpm) 48,000
Max. RCF (x g) 118,000
Max. revolving radius: Rmax (mm) 66
Rotor weight (kg) 31
Rotor material Titanium alloy
RCF (x g) at outside radius of center body 98,000 37,600 48.300
Radius of center body: Rmin (mm) 55 21 55 27
Core length (mm) 761
Core material Noryl® Titanium Noryl® (Polyphenilen Oxide) Registered
by General Electric
Capacity (l) 3.2 8 3.2
Preclarifier 0.3
7.7
Max. flow rate (l/hr) 50 - 50
O-ring material EPDM (Ethylene-propylene rubber, USP Class VI compliant)
Applications Density gradient centrifugation of virus, liposome, etc. Batch centrifugation of HBsAg etc. Density gradient centrifugation of virus, liposome, etc. The preclarifier core permits separation of cells and fractions. Rough separation of large-volume samples for pelleting.

Notes:
  1. The C40CT4 rotor assembly, using Core (H) and Titanium core (H), is designed with 30% lower flow restriction compared to the C40CT3 rotor assembly using the same cores. Which allows for a higher flow rate for viscous samples.
  2. The Core (E) is not for continuous flow but for batch centrifuge operation.
  3. The shape and size are the same as Core (H).
  4. The shape and size are the same as Core (D).

page top

Rotors for CC40S
Rotor model C40CTS4 *1
C40CTS3
Type of cores Core (AH)
(High Flow Type)
Core (A) Ti Core (B)
External appearance of cores
With preclarifier core
Max. speed (rpm) 40,000
Max. RCF (x g) 118,000
Max. revolving radius: Rmax (mm) 66
Rotor weight (kg) 15
Rotor material Titanium alloy
RCF (x g) at outside radius of center body 98,000
Radius of center body: Rmin (mm) 55
Core length (mm) 382
Core material Noryl® (Polyphenilen Oxide) Registered
by General Electric
Titanium
Capacity (l) 1.6 1.6
Preclarifier 0.15
Max. flow rate (l/hr) 50
O-ring material EPDM (Ethylene-propylene rubber, USP Class VI compliant)
Applications Density gradient centrifugation of virus, liposome, etc. (For high viscocity sample) Density gradient centrifugation of virus, liposome, etc. Density gradient centrifugation of virus, liposome, etc. The preclarifier core permits separation of cells and fractions.

Note: * The C40CTS4 rotor assembly and Core (AH) is designed with 30% lower flow restriction compared to the C40CTS3 rotor assembly and Core(H). Which allows for a higher flow rate for viscous samples.

Features
Specifications

page top

Terms of Use, etc.
Hitachi Koki Co., Ltd. Copyright(C) 2000-2006 Hitachi Koki Co., Ltd All rights reserved .