Hitachi Koki HITACHI
HomePower ToolsLife scienceFinancialCorprate ProfileContact us Japanese

Life-science > Products > High-Capacity Refrigerated Centrifuge

Products

High-Capacity Refrigerated Centrifuge


CR7
The brushless motor emits no dust, making this unit ideal for use in clean hospital environments and for blood component separation in blood centers.
Powerful refrigeration and fine control maintain constant sample temperatures.

*The photo shows a CR7 with log data printer (option) connected.

Specifications
Rotors for CR7

Features

High-capacity carbon fiber rotor

  • 1,000 ml x 6 bottles
  • Max. 7,000 rpm (10,700 x g)
  • To be used exclusively with a wide-mouth 1-liter bottle
    1000PP bottle (WM)(P/N: S307700A, 6 pcs/box)
    1000PC bottle (WM)(P/N: S308357A, 6 pcs/box)
    1000PP cap (WM)(P/N: S410870A, 2 pcs/box)
  • For high-volume separation of bacteria and organelle from their culture liquid and for rough separation in the manufacturing process of vaccines

High-capacity swing rotor

  • 400 ml MAP bag x 12 units can be processed at a time.
  • Max. 4,200 rpm (5,030 x g)
  • For blood component separation use at blood centers, etc.

Advanced technology

  • Memory operation
    Stores 30 sets of operating parameters.

  • Net run time input
    Because the timer starts simultaneously with attainment of the selected speed, net operation time can be realized (patent pending).

  • Convenient functions are standard features
    • RCF calculation
    • Memory lock function
    • g• sec. calculation
    • Step operation input

  • Numerous safety devices and self-diagnosis functions are standard
    • Door opening/closing detection
    • Imbalance detection
    • Overspeed detection
    • Rotor setting abnormality detection
    • Temperature control abnormality detection
    • Power source abnormality detection
    • Motor rotation abnormality detection

GLP/GMP-compatible operation record control

  • If a log data printer is connected, operating conditions and operating results can be printed out. This function is convenient for the control of GLP/GMP-related manufacturing lines and for operation record management at the inspection room.

  • CR7-type log data printer (option)
    This printer is designed to print out speed, time, humidity and 11 other operating conditions for start, stop and alarm events.

    *In the case of on-site installation, there is a charge for upgrade labor and travel expenses.

Specifications

Model
CR7
Maximum speed (rpm)
7,000
Maximum RCF (x g)
10,700 (R7AF3 rotor)
Maximum capacity (tubes x ml)
1,000 x 6 (R7AF3 rotor or R5S2 rotor)
Maximum speed at maximum-capacity rotor and maximum RCF
7,000 rpm, 10,700 x g
Drive motor
High-torque brushless high-frequency motor
Speed control range (rpm)
300 to 7,000
Temperature range adjustment (°C)
-20 to +40
Timer
One second to 99 minutes and 59 seconds, with a HOLD function (one minute to 99 hours and 59 minutes optional)
Refrigerator capacity/refrigerant
1.5 kW/R404a
Function Memory-based programmed operation Nine programmed operations possible
Acceleration/deceleration time variable Nine-stage variable acceleration, nine-stage braked deceleration, plus free deceleration
RCF calculation
RCF integrator
RTC (real-time control)* N/A
Rotor stoppage buzzer
Imbalance detection device
Automatic rotor temperature compensation
Self-diagnosis
Automatic rotor identification N/A
Continuous flow rotor N/A
Power requirements
Single phase, AC 220/230/240 V ±10%, 30A, 50/60Hz
Dimensions (mm)
730 (W) x 910 (D) x 1,150 (H)
Height to the controller (mm) 870
Standard accessories Instruction manual, rotor cleaning bar, silicone grease, fuse (0.4A) surge type, fuse (2A) and fuse (25A)
Weight (kg)
300

Note:*CR7 has no RTC functions.

Cautions during installation

  1. To optimize the heat exchange efficiency of the refrigerator unit, install in a space according to the specifications.
  2. Install on a stable floor.
  3. Avoid installation near equipment that could be a heat source, or in direct sunlight.
  4. An extension cord should not be used.
  5. The heat radiation rate of the centrifuge is about 2,300 kcal/h.
  6. As a safety precaution, ground the equipment properly according to applicable electrical codes.

page top