 High-Capacity Refrigerated Centrifuge

CR7 |
The brushless motor emits no dust, making this unit ideal for use in clean hospital environments and for blood component separation in blood centers.
Powerful refrigeration and fine control maintain constant sample temperatures.
*The photo shows a CR7 with log data printer (option) connected.
|
Features
High-capacity carbon fiber rotor
- 1,000 ml x 6 bottles
- Max. 7,000 rpm (10,700 x g)
- To be used exclusively with a wide-mouth 1-liter bottle
1000PP bottle (WM)(P/N: S307700A, 6 pcs/box)
1000PC bottle (WM)(P/N: S308357A, 6 pcs/box)
1000PP cap (WM)(P/N: S410870A, 2 pcs/box)
- For high-volume separation of bacteria and organelle from their culture liquid and for rough separation in the manufacturing process of vaccines
High-capacity swing rotor
- 400 ml MAP bag x 12 units can be processed at a time.
- Max. 4,200 rpm (5,030 x g)
- For blood component separation use at blood centers, etc.
Advanced technology
- Memory operation
Stores 30 sets of operating parameters.
- Net run time input
Because the timer starts simultaneously with attainment of the selected speed, net operation time can be realized (patent pending).
- Convenient functions are standard features
• RCF calculation
• Memory lock function
• g• sec. calculation
• Step operation input
- Numerous safety devices and self-diagnosis functions are standard
• Door opening/closing detection
• Imbalance detection
• Overspeed detection
• Rotor setting abnormality detection
• Temperature control abnormality detection
• Power source abnormality detection
• Motor rotation abnormality detection
GLP/GMP-compatible operation record control
- If a log data printer is connected, operating conditions and operating results can be printed out. This function is convenient for the control of GLP/GMP-related manufacturing lines and for operation record management at the inspection room.
CR7-type log data printer (option)
This printer is designed to print out speed, time, humidity and 11 other operating conditions for start, stop and alarm events.
*In the case of on-site installation, there is a charge for upgrade labor and travel expenses.
|
 Specifications
| Maximum speed (rpm) |
7,000
|
| Maximum RCF (x g) |
10,700 (R7AF3 rotor)
|
| Maximum capacity (tubes x ml) |
1,000 x 6 (R7AF3 rotor or R5S2 rotor)
|
| Maximum speed at maximum-capacity rotor and maximum RCF |
7,000 rpm, 10,700 x g
|
| Drive motor |
High-torque brushless high-frequency motor
|
| Speed control range (rpm) |
300 to 7,000
|
| Temperature range adjustment (°C) |
-20 to +40
|
| Timer |
One second to 99 minutes and 59 seconds, with a HOLD function (one minute to 99 hours and 59 minutes optional)
|
| Refrigerator capacity/refrigerant |
1.5 kW/R404a
|
| Function |
Memory-based programmed operation |
Nine programmed operations possible |
| Acceleration/deceleration time variable |
Nine-stage variable acceleration, nine-stage braked deceleration, plus free deceleration |
| RCF calculation |
 |
| RCF integrator |
 |
| RTC (real-time control)* |
N/A |
| Rotor stoppage buzzer |
 |
| Imbalance detection device |
 |
| Automatic rotor temperature compensation |
 |
| Self-diagnosis |
 |
| Automatic rotor identification |
N/A |
| Continuous flow rotor |
N/A |
| Power requirements |
Single phase, AC 220/230/240 V ±10%, 30A, 50/60Hz
|
| Dimensions (mm) |
730 (W) x 910 (D) x 1,150 (H)
|
| Height to the controller (mm) |
870 |
| Standard accessories |
Instruction manual, rotor cleaning bar, silicone grease, fuse (0.4A) surge type, fuse (2A) and fuse (25A) |
| Weight (kg) |
300
|
| Note:*CR7 has no RTC functions. Cautions during installation
- To optimize the heat exchange efficiency of the refrigerator unit, install in a space according to the specifications.
- Install on a stable floor.
- Avoid installation near equipment that could be a heat source, or in direct sunlight.
- An extension cord should not be used.
- The heat radiation rate of the centrifuge is about 2,300 kcal/h.
- As a safety precaution, ground the equipment properly according to applicable electrical codes.
|
|
|

|